Measurement Accuracy |
|
Voltage DC |
30.000 V |
0.01%+ 2 mV (upper display) |
20.000 V |
0.01%+ 2 mV (lower display) |
90.00 mV |
0.01%+ 10μV |
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
Current DC |
24.000 mA |
0.01%+ 2 mA |
Resistance |
0.0 to 4000.0 Ω |
+/- 0.015% |
Frequency |
2.0 to 1000.0 CPM |
0.05 % + 0.1 cpm |
1.0 to 1000.0 Hz |
0.05 % + 0.1 Hz |
1.00 to 15.00 kHz |
0.05 % + 0.01 kHz |
Sensitivity |
1 V peak-to-peak-minimum |
Pressure |
Accuracy |
from 0.025% of range using any of 29 pressure modules. |
|
(for detailed specifications refer to pressure modules in options and accessories) |
|
Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
Source Accuracy |
Voltage DC |
100.00 mV |
0.01%+10 μV |
20.000 V |
0.01%+2 mV |
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
Current DC |
24.000 mA (Source) |
0.01%+ 2 μA |
24.000 mA (Simulate) |
0.01%+ 2 μA |
Resistance |
Range, Accuracy (Percent of reading + floor) and Excitation Current |
5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5mA |
|
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3mA |
|
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA |
|
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA |
Frequency |
2.0 to 1000.0 CPM |
0.05% |
1.0 to 1000.0 Hz |
0.05% |
1.0 to 10.0 kHz |
0.25% |
10.0 to 15.0 kHz |
0.5% |
Waveform: |
5 V p-p squarewave, -0.1 V offset |
RTDs and Thermocouples |
Measure and Source Accuracy: |
Cu10 |
1.8 °C |
NI-120 |
0.15 °C |
PT-100 (385) |
0.45 °C |
PT-100 (392) |
0.35 °C |
PT-100 (391-JIS) |
0.35 °C |
PT-200 (385) |
0.95 °C |
PT-500 (385) |
0.55 °C |
PT-1000 (385) |
0.35 °C |
RTD Resolution: |
0.01 C (0.01 F) |
Cu10.01 C |
0.1 C (0.1 F) |
J: |
0.4 °C |
K: |
0.5 °C |
T: |
0.4 °C |
E: |
0.4 °C |
R: |
1.4 °C |
S: |
1.4 °C |
B: |
1.4 °C |
L: |
0.45 °C |
U: |
0.45 °C |
N: |
0.6 °C |
XK: |
0.4 °C |
BP: |
1.1 °C |
TC Resolution |
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F |
Specifications |
Ramp Functions |
Source functions: |
Voltage, current, resistance, frequency, temperature |
Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
Loop Power Function |
Voltage: |
24 V |
Accuracy: |
10% |
Maximum current: |
22 mA, short circuit protected |
Step Functions |
Source functions: |
Voltage, current, resistance, frequency, temperature |
Steps |
25% of range, 100% of range |
Environmental Specifications |
Operating Temperature |
|
-10 °C to +55 °C |
Storage Temperature |
|
-20 °C to +71 °C |
Operating Altitude |
|
3000 m |
Safety Specifications |
Agency Approvals |
|
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
Mechanical & General Specifications |
Size |
|
200mm x 96mm x 47mm (7.9 in x 3.8 in x 1.9 in) |
Weight |
|
650 g (23 oz.) |
Batteries |
|
4 AA alkaline batteries |
Battery Replacement |
|
Separate battery compartment, accessible without breaking calibration seall |
Side port connections |
|
Pressure module connector, also used for remote real-time programming |